|
Suggestions for Improved Reliability
in Thermal Analysis of Cast Irons
1. Liquid Iron Sampling - uniformity
& consistency are mandatory
- Make sure the sample ladle is free of
iron and slag and is as hot as practical.
- Don't use graphite or clay-graphite
ladles for sampling since they tend to produce some inoculating
effect.
- Make sure the iron being sampled is as
close to the same temperature as possible from sample to sample.
(Note that this temperature will vary depending on whether it is a
furnace or ladle sample.)
- For electric furnace samples, best
results are usually obtained once a temperature of ~2700oF
has been reached in the furnace.
- If sample ladle is filled from another
ladle, we recommend filling the sample ladle, emptying it and
refilling before pouring your sample.
- Make sure the cup stand is kept as clean
as possible, free from resin build-up and spilled iron, out of any
cold drafts and at as constant a temperature as possible.
- Make sure the cup and stand are level,
not tipped, and that the cup is poured completely full.
- If frequent samples are poured (more
than ~4-5/hr.), use a second stand and alternate between stands to
prevent overheating of the stand.
- Good sampling practice will result in
consistent maximum temperature at least 50oF to 100oF
above the liquidus temperature.
- When pouring final iron samples, allow
at least two minutes after inoculant addition before pouring your
sample.
- Remove the sample from the cup stand as
soon as analysis is completed to minimize resin build-up and to permit
the stand to cool prior to the next test.
- Regularly clean the cup stand and
contacts with a wire brush or other means.
2. Equipment Calibration
- from the cup all the way back to the
instrument - extremely important!! (Be aware of variations between
IPTS thermocouple standards. Cups for the N. American foundry
industry are 1948 IPTS. Do you know whether your instrument is
being calibrated per 1948, 1968, or 1990 IPTS?)
Table showing errors between IPTS 68 and
IPTS 48 and also the temperature differences which exist if IPTS 48
thermocouples are used with IPTS 68 instruments.
(From 9/23/81 memo from L.R. Jones/E-N to
W.F. Shaw)
| Temp
oF |
Instrument
68/TC 48 |
| IPTS 68 |
IPTS 48 |
Temp oF |
ToF |
| 2000 |
1997.3 |
1995.5 |
4.5 |
| 2100 |
2097.1 |
2095.4 |
4.6 |
| 2200 |
2196.9 |
2195.1 |
4.9 |
| 2300 |
2296.8 |
2294.9 |
5.1 |
| 2400 |
2396.7 |
2394.0 |
6.0 |
| 2500 |
2496.5 |
2492.5 |
7.5 |
3. Cup purchasing, storage &
monitoring
- Purchase cups in large quantities
(obviously based on your usage) with the specification that all cups
in each order be from the same manufacturing lot.
- Store cups in as warn and dry an
atmosphere as possible. If stored in a cold warehouse, make sure
cups are brought up to room temperature well in advance of their
usage.
- Before using a new lot number, compare
TA curves from current vs. new lot of cups to ensure that no
significant differences occur.
- Include the supplier's lot number on
your melt records & note when a change in lot number occurs.
|